Acoustical Signals - Handpiece
An acoustical signal is emitted
• every time a button is pressed,
• every time the light is turned ON,
• 1 time after 5 sec exposure time, 2 times after 10 sec,
3 times after 15 sec.
Exception: in continuous mode; an acoustical signal
is emitted every 10 seconds.
Two acoustical signals are emitted
• every time the sleep mode is terminated by pressing
the START button,
• every time the light is turned OFF.
A 2-sec error signal is emitted, if
• the handpiece over-heats,
• the battery lacks sufficient charge.
The acoustical signals from the handpiece can be turned
off (except for the 2-sec error signal). Follow these
instructions to turn them off. Put the handpiece in sleep
mode, e. g., by setting it in the charger. Take the device
from the charger; press first the TIME button, then the
START button. This changes the sleep mode of the
hand piece back to active and the status “Acoustical
signals activated” to “Acoustical signals deactivated”.
The acoustical signals can be reactivated by following
the same procedure.
Troubleshooting
Error
Cause
왘
Solution
The charge level
display on the hand-
piece glows red
steadily.
The power level
display of the hand-
piece flashes red.
The ongoing exposure
is interrupted (light off
signal is emitted) fol -
lowed by a 2-sec error
signal; the handpiece
switches to “sleep”
mode and resists
further activation.
The remaining battery charge
is adequate only for typical 5
10-sec exposure cycles.
왘
Place the handpiece in the
charger and re-charge the
battery.
The battery lacks sufficient
charge.
왘
Place the handpiece in the
charger and re-charge the
battery.
Error
Cause
왘
Solution
The power level
display of the hand-
piece flashes red
while the handpiece
is in the charger.
The handpiece has not
been used for a long
time and now it cannot
be turned on.
The light emission
does not start when
the START button is
pressed; an error
signal sounds for
2 sec.
During light emission
in continuous mode,
an error signal sounds
for 2 sec, the emission
is stopped, and the
handpiece changes to
sleep mode.
The LED showing the
operating status in
the charger glows red
steadily.
Charging problem.
The battery is defective or at
the end of its useful life.
왘
Replace the battery.
There is not enough charge
in the battery to turn on the
handpiece.
왘
Place the handpiece in the
charger and re-charge the
battery.
The information about inter-
mittent operation under the
header “Technical Data -
Handpiece” has not been
observed.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
왘
Allow the handpiece to
cool for 3 minutes, then
start the next exposure by
pressing the START button.
The information about inter-
mittent operation under the
header “Technical Data -
Handpiece” has not been
observed.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
왘
Allow the handpiece to
cool for 3 minutes, then
start the next exposure by
pressing the START button.
The charge contact pins in
the handpiece or the charger
are wet.
왘
Dry the charge contact
pins. Be careful not to
bend the flexible charge
contact pins in the charger.
Содержание Elipar DeepCure-S
Страница 77: ...76 LED LED LED 1 5...
Страница 80: ...79 el 3M ESPE Sleep Mode 5 START 1 5 2 10 3 15 10 START 2 2 sleep mode TIME START...
Страница 81: ...80 4 2 5 10 4 START 2 2 LED 3 START 3 START...
Страница 82: ...81 el 4 LED LED Elipar DeepCure S Elipar DeepCure S 3M ESPE...
Страница 83: ...82 1 5 LED 3M Deutschland GmbH pH 5 5 pH 8 5 134 C 273 F 500 2 10 K K Cidezyme Enzol Johnson Johnson 10...