Guidance and manufacturer’s declaration – electromagnetic immunity
The Elipar DeepCure-S Dental Light Unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the Elipar DeepCure-S Dental Light Unit should assure that it is used in such an
environment.
IEC 60601
Electromagnetic environment -
Immunity test
Compliance level
test level
guidance
Electrostatic
± 6 kV contact
± 6 kV contact
Floors should be wood, concrete or
discharge (ESD)
ceramic tile. If floors are covered with
± 8 kV air
± 8 kV air
synthetic material, the relative humidity
IEC 61000-4-2
should be at least 30 %.
Electrostatic
± 2 kV for power
± 2 kV for power
Mains power quality should be that
transient / burst
supply lines
supply lines
of a typical commercial or hospital
environment.
IEC 61000-4-4
± 1 kV for input/output
± 1 kV for input/output
lines
lines
Surge
± 1 kV differential
± 1 kV differential
Mains power quality should be that
mode
mode
of a typical commercial or hospital
IEC 61000-4-5
environment.
± 2 kV common mode
± 2 kV common mode
Voltage dips, short
< 5 % U
T
< 5 % U
T
Mains power quality should be that of
interruptions and
(> 95 % dip in U
T
)
(> 95 % dip in U
T
)
a typical commercial or hospital
voltage variations
for 0,5 cycle
for 0,5 cycle
environment. If the user of the
on power supply
Elipar DeepCure-S Dental Light Unit
input lines
40 % U
T
40 % U
T
requires continued operation during
(60 % dip in U
T
)
(60 % dip in U
T
)
power mains interruptions, it is
IEC 61000-4-11
for 5 cycles
for 5 cycles
recommended that the Elipar
DeepCure-S Dental Light Unit be
70 % U
T
70 % U
T
powered from an uninterruptible power
(30 % dip in U
T
)
(30 % dip in U
T
)
supply or a battery.
for 25 cycles
for 25 cycles
< 5 % U
T
< 5 % U
T
(>95 % dip in U
T
)
(> 95 % dip in U
T
)
for 5 sec
for 5 sec
Power frequency
3 A/m
3 A/m
Power frequency magnetic fields
(50/60 Hz)
should be at levels characteristic of a
magnetic field
typical location in a typical commercial
or hospital environment.
IEC 61000-4-8
NOTE
U
T
is the a. c. mains voltage prior to application of the test level.
Содержание Elipar DeepCure-S
Страница 77: ...76 LED LED LED 1 5...
Страница 80: ...79 el 3M ESPE Sleep Mode 5 START 1 5 2 10 3 15 10 START 2 2 sleep mode TIME START...
Страница 81: ...80 4 2 5 10 4 START 2 2 LED 3 START 3 START...
Страница 82: ...81 el 4 LED LED Elipar DeepCure S Elipar DeepCure S 3M ESPE...
Страница 83: ...82 1 5 LED 3M Deutschland GmbH pH 5 5 pH 8 5 134 C 273 F 500 2 10 K K Cidezyme Enzol Johnson Johnson 10...