3
switches to sleep mode once it is placed in the charger or
if left unused for approx. 5 minutes outside the charger.
Do not discard these Instructions for the duration of
product use.
Fields of Application
• Polymerization of light-curing dental materials with
photo initiator for the wavelength range 430-480 nm.
- Though the majority of light-curing dental materials
are responsive in this range of wavelengths,
you may wish to contact the manufacturer of the
material in question.
Technical Data
Charger
Operating voltage:
100-127 V 50/60 Hz
230 V 50/60 Hz (see rating
plate for factory-set voltage)
Power input:
20 VA
Dimensions:
Length: 170 mm
Width:
95 mm
Height:
50 mm
Weight:
650 g
Classification:
Protection class II,
Handpiece
Power supply:
lithium-ion battery, nominal
voltage 3.7 V
Utilizable wavelength
range:
430-480 nm
Wavelength peak:
444-452 nm
Light intensity (between
400 and 515 nm):
1470 mW/cm
2
-10%/+20%
(independent of battery power
level)
Light emission area:
60-65 mm
2
(optically active)
Intermittent operation: The device has been designed
solely for short-term operation.
Typical operating time at room
temperature (23° C): 7 min,
at 40° C ambient temperature:
1 min on, 15 min off (cooling-off
period)
Total exposure time with new,
fully charged battery:
typically 120 min
Dimensions:
Diameter: 28 mm
Length:
270 mm
Weight:
250 g (incl. light guide)
Typ B
In comparison with conventional light-curing devices,
Elipar DeepCure-S features excellent beam collimation
and a uniform beam profile, directing more of the light
energy to the restoration being polymerized and
producing a deep, uniform, and complete curing of the
restoration.
The light source is a high-performance light diode (LED).
The beam emerging from the device covers the light
wavelength range of 430 to 480 nm relevant, for
instance, for camphor quinone products and is suitable
for use with the majority of light-curing dental materials,
including materials for fillings, liners, core build-ups,
fissure sealings, temporary restorations, and cements
for indirect restorations.
See the manufacturer’s information for the exposure
time required for the specific dental material.
Settable exposure times:
• 5, 10, 15, 20 sec
• Continuous mode (120 sec)
• Tack-cure mode (1 sec)
The charger is equipped with an integrated light intensity
meter for the Elipar DeepCure-S.
The unit is shipped with a light guide with 10 mm
diameter. It is not permissible to use the light guides of
other units.
The handpiece is equipped with a sleep mode to mini-
mize the unit’s energy consumption. The handpiece
ENGLISH
en
Light guide
Handpiece
Charger
Power cord
Glare Shield
Содержание Elipar DeepCure-S
Страница 77: ...76 LED LED LED 1 5...
Страница 80: ...79 el 3M ESPE Sleep Mode 5 START 1 5 2 10 3 15 10 START 2 2 sleep mode TIME START...
Страница 81: ...80 4 2 5 10 4 START 2 2 LED 3 START 3 START...
Страница 82: ...81 el 4 LED LED Elipar DeepCure S Elipar DeepCure S 3M ESPE...
Страница 83: ...82 1 5 LED 3M Deutschland GmbH pH 5 5 pH 8 5 134 C 273 F 500 2 10 K K Cidezyme Enzol Johnson Johnson 10...