![3DHISTECH P1000 Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/3dhistech/p1000/p1000_service-manual_442157093.webp)
P1000 SERVICE SOFTWARE MANUAL
16. OBJECTIVE CHANGER UNIT
92
SERVICE GUIDE
16. Objective Changer Unit
In this setup window, the objective, aperture and ND filter positions can be synchronized. For proper operation, synchronize the settings of the
objectives first.
Note!
The Objective change position value is identical with the objective position already set in
1.
Click
Objective changer
option in the Sub-assembly column on the left and, on Validation page, click the
Focus motor position
button. Send the
actuator to this position by clicking
Go to II
.
2.
In Objective changer unit settings, click Objective change position. Save the value.