P1000 SERVICE SOFTWARE MANUAL
10. ND FILTER
30
SERVICE GUIDE
3.
The high-
level setup window is the same as the aperture’s. Create three positions and select position one (filtered position).
4.
Click the
Slice motor position
button to activate the motor control panel. Click the
Home
button.
5.
Step the disk in positive direction until the filter and the objective behind are aligned (you can light the unit with a lamp from the direction of the
condenser). When done, click
on the side panel to set this value and to the
Slice motor position
cell.
6.
Repeat steps from 10.3 to 10.5 to find the other two filter positions. Click
to set the values.
7.
After defining all three values, start the validation process. Set
Test count
to the intended value or set 1 and validate the component.