P1000 SERVICE SOFTWARE MANUAL
13. STAGE
SERVICE GUIDE
53
4.
Repeat steps in 13.2 and 13.3 in Y direction, too. Note that the end stop pins in Y are not visible. In this case, the sound of the collision and the
changing value in the
Pos field
help find the end positions. At the end position, both opening arms must have minimal room for maneuver. If any
of the arms is in contact, you must place a shim for a proper setting.
5.
In Z direction, the unit does not stop at the end position. It tries to reach the adjusted position and push the lifting frame forward. Do not allow
the unit to keep trying long, press
Stop
to cancel.
5.1.
To find the positive end position, place a slide with a cover slip into the Stage under the objective. The distance between the objective
(ideally, Zeiss 20X) and the slide must be 0,2mm. Make this adjustment lying a 0.2mm gauge plate on the slide and stepping the focus unit in
positive direction until touching the slide. Fix the adjustable stop end screw with a counter nut. After fixing, verify the distance again using
the gauge plate!
Lifting frame
Encoder (Renishaw)
Stop end screw
Piezo motor
Содержание P1000
Страница 1: ...P1000 SERVICE SOFTWARE MANUAL SERVICE GUIDE JANUARY 13 2020 3DHISTECH Cég címe ...
Страница 62: ...P1000 SERVICE SOFTWARE MANUAL 14 SLIDE LOADING UNITS SERVICE GUIDE 61 ...
Страница 72: ...P1000 SERVICE SOFTWARE MANUAL 14 SLIDE LOADING UNITS SERVICE GUIDE 71 Back end pin Front end pin ...
Страница 88: ...P1000 SERVICE SOFTWARE MANUAL 14 SLIDE LOADING UNITS SERVICE GUIDE 87 7 3 Safety container position ...