14. SLIDE LOADING UNITS
86
SERVICE GUIDE
To set the decision limit, click the icon next to the
Slide edge
label at the component control panel and then click the
Turn on
button below
the icon. The display of the sensor is activated. Navigate the Arm with the
Approach the slide
button to a position where the magazine does
not include a slide. Push the
Set
button on the sensor
.
Navigate the Arm to slide position and push Set again. The sensor calculates an average
decision threshold from the two values and signs if there is a slide in the actual position.
If the detection of all inserted magazines is successful, click the
Sensor is working
button.
7.2.
Fine tuning the Slide out sensor
Turn on the sensor on the control panel.
You can navigate the Arm to each column in
Sensors column setup
. Check if the laser lights on the prism. If it does not, the bottom prisms
must be aligned so that the laser dot lights the center of the prism. When you are ready, click the tick under the intended column.