![3DHISTECH P1000 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/3dhistech/p1000/p1000_service-manual_442157035.webp)
P1000 SERVICE SOFTWARE MANUAL
11. OBJECTIVE CHANGER
34
SERVICE GUIDE
7.
Find the safety Z position to change the objective safely using the control panel of the motor. Place a flat slide on the top of the related objective
mount as illustrated below. Do not place the slide on the retaining ring.
The adjustment is correct if you move the unit in Z until the slide is parallel with the top surface of the two Objective mounts. Click
on the side
panel to set this value.
Objective
retaining ring
Flat metal slide