Page 72 User manual 5448X000 Version 1 June-06
Wavelength calibration (wavelength adj)
Press
(8)
key to carry out wavelength
calibration function, as shown in picture
52.
Quitting utilities mode
Press
Return
key (4.3e) to quit utilities
mode and return to the instrument main
menu.
14:57
Waiting ...
Utilities
1. Lamp Chang Over WL: 359.0
2. Bandwidth: 2.0
3. W Lamp: On
4. D2 Lamp: On
5. Control Mode: PC
6. Set Time
7. Correcting Dark Cur.
8. Correcting Wave Length
Picture 52
6. SECURITY
The spectrophotometer must be used by previously qualified staff that knows how the
equipment works thanks to the user manual.
You should put the spectrophotometer in a horizontal plane schart chart, having a
safety area of at least 30 cm per side.
Do not place the spectrophotometer near any warm supply (burners, blowlamps, etc),
nor expose it directly to the sun. Avoid vibrations, dust and dry environments.
When you are not using the spectrophotometer for a long period of time please make
sure it is unplugged in order to avoid possible accidents.
It is essential to have the equipment switched off and unplugged from the net before
cleaning, checking components or replacing any piece (e.g. replacement of a fuse).
Never try to repair the spectrophotometer by yourself, since you will lose the warranty
and may provoke damages to the general operating system or the electrical installation,
as well as injuries to the people that usually handle the equipment (burns, hurts…).
Try not to spill any liquid on the control panel, though it is properly insulated. In case
you have any doubts do please immediately unplug the spectrophotometer (see
Maintenance
).
Made under the European regulations for electrical security, electromagnetic compati-
bility and security on machines.
Version 1 June-06
User manual 5448X000 Page 41
INDEX OF CONTENTS
1. USES OF THE INSTRUMENT
2. DESCRIPTION
3. TECHNICAL SPECIFICATIONS
4. INSTALLATION / SETTING UP
5. FUNCTIONS
6. SECURITY
7. MAINTENANCE
8. TROUBLESHOOTING
ANNEX I: CE CERTIFICATE
41
41
44
45
48
72
73
75
77
1. USES OF THE INSTRUMENT
2. DESCRIPTION
Zuzi UV Visible spectrophotometers series 4488/4481 are versatile instruments availa-
ble for a wide range of uses. This is posible thanks to functions as: absorbance, transmit-
tance, concentration calculus, quantitative readings from straight line standards or DNA
measuring.
Among the variety of fields of use where thespectrophotometers can be used we stand
out : petrochemical and biochemical industries, mining industry, agricultural engineering,
hospitals, environmental conservation and scientific R&D (Research and Development)
laboratories.
1
2
3
1.1. Sample compartment
1.2. LCD display
1.3. Keyboard
Picture 1 - Front part
Sample compartment
Provided with 8-cell holder for samples.
LCD display
The 320×240 dots high resolution LCD. In order to adjust the brightness you should
insert the tip of a plastic screwdriver into a hole located on the left corner of the keyboard.
On display asre shown image information, parameters, data and spectrograms.
Keyboard
27 soft-touch keys on the keyboard are used for instrument control and operation.
ENGLISH
ENGLISH