Specifications
A-20
www.zoll.com
9650-0806-01 Rev. D
Pulse Oximeter
Range:
Oxygen Saturation (% SpO
2
)
1% - 100%
Carboxyhemoglobin Saturation (% SpCO)
0% - 99%
Methemoglobin Saturation (% SpMet)
0% - 99%
Pulse Rate (bpm)
25 - 240 beats per minute
Accuracy:
Oxygen Saturation (% SpO
2
) - During No Motion Conditions
Adults, Pediatrics
1
70% - 100%, ±2 digits 0% - 69%, unspecified
Neonates
2
70% - 100%, ±3 digits 0% - 69%, unspecified
Oxygen Saturation (% SpO
2
) - During Motion Conditions
3
Adults, Pediatrics
70% - 100%, ±3 digits 0% - 69%, unspecified
Neonates
70% - 100%, ±3 digits 0% - 69%, unspecified
Oxygen Saturation (% SpO
2
) - During Low Perfusion Conditions
4
Adults, Pediatrics
70% - 100%, ±2 digits
Neonates
70% - 100%, ±3 digits
Carboxyhemoglobin Saturation (% SpCO)
5
1% - 40% ±3 digits
Methemoglobin Saturation (% SpMet)
5
1% - 15% ±1 digit
Pulse Rate (bpm) - During No Motion Conditions
1
Adults, Pediatrics, Neonates
25 - 240 ±3 digits
Pulse Rate (bpm) - During Motion Conditions
3
Adults, Pediatrics, Neonates
25 - 240 ±5 digits
Resolution:
SpO
2
: 1%
SpCO: 1%
SpMet: 0.1% for range up to 9.9%, 1% for range 10 - 99%
Pulse rate: 1 bpm (beats per minute)
Alarm Limits:
On/Off displayed on monitor. User selectable.
SpO
2
: High 72 - 100% saturation, Low 70 - 98% saturation
SpCO: High 2 - 100% saturation, Low 0 - 99% saturation
SpMet: High 1 - 100% saturation, Low 0 - 99% saturation
Pulse Rate: High 60 - 235 beats per minute, Low 20 - 100 beats per minute
SpO
2
Wavelength for LNCS Sensors:
Nominal Red LED Wavelength: 660 nanometers
Nominal Infrared LED Wavelength: 905 nanometers
Energies (Radiant Power) of light for
LNCS Sensors at 50 mA pulsed:
15 mW
SpO
2
Wavelength for Rainbow Sensors:
The Rainbow sensors use 8 different LEDs with wavelengths of 610 - 905
nanometers
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...