Verification Prior to Use
Before each use with the Propaq MD, verify the proper operation of the ZOLL internal paddles
using the following procedure. This procedure requires a second person if you are using
internal handles without a Discharge button.
Warning!
When performing internal paddle verification, keep hands away from the electrode
plates while pressing the SHOCK button.
1. Inspect the connector contact sockets for damage or corrosion. If you observe damage or
corrosion in the connector contact sockets, remove the handle set from use.
2.
Connect the Autoclavable Internal Handles to the Propaq MD unit. Use the Lead quick access
key
to verify that the Propaq MD unit correctly identifies the Internal Handle and
Electrode set by displaying
Int. Pdls
.
3.
Before charging the defibrillator, press the
Discharge
button on the handle set (if present) and
verify that there is an audible click and that the button springs back upon release. Verify that
the defibrillator window displays the message
APPLY PADDLES TO PATIENT
. This message
verifies that the
Discharge
button located on the right handle is operating correctly.
4.
Press the Select Energy arrows (located on the front panel of the Propaq MD unit) up or down
to select 30 Joules.
5.
Press the
Charge
button on the defibrillator front panel to charge the unit to the selected
energy level. Wait for the READY tone.
6.
Press the electrode plate surfaces firmly together and away from any person or object.
7.
Discharge the energy in the following manner.
•
For internal handles with a Discharge button:
Press and hold the
Discharge
button on the apex handle to deliver the test energy to the
electrodes.
•
For internal handles without a Discharge button:
Have a second person press and hold
on the defibrillator front panel to deliver the test
energy to the electrodes.
The Propaq MD device unit discharges and displays the message
DEFIB SHORT TEST
PASSED
.
I, II,
III...
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...