Configuring NIBP Alarms and Settings
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
8-9
In order to accurately measure systolic pressure, the CIT pressure must be high enough to
occlude the underlying artery. However, setting the CIT pressure too high may unnecessarily
increase the reading determination time and patient discomfort.
After each NIBP measurement, the Propaq MD unit adjusts the cuff inflation pressure to
optimize the next NIBP measurement.
Note:
The maximum cuff inflation pressure for neonates is 153 mmHg.
Warning!
Before using the Propaq MD unit to monitor a new patient, power down the unit for at
least 2 minutes to reset all patient parameters and eliminate all adjustments made for
the previous patient.
Configuring NIBP Alarms and Settings
The last step in preparing to perform NIBP measurements is to ensure that the necessary alarms
are enabled (or disabled), that alarm limits are appropriate, and that the NIBP settings are
correct.
Enabling/Disabling NIBP Alarms and Setting Alarm Limits
When enabled, the Propaq MD unit sounds alarms whenever measurements are outside set
limits for the following:
•
High and Low Systolic Pressure
•
High and Low Diastolic Pressure
•
High and Low Mean Arterial Pressure (MAP)
To enable (or disable) NIBP alarms and set upper and lower alarm limits, you can do so through
the Alarms
quick access key (
) or the
NIBP Parameter Control Panel
.
To configure NIBP alarm limits through the Alarms quick access key:
1. Press the More quick access key (
).
2.
Press the Alarms quick access key (
).
3.
Press the
Limits
quick access key (
).
4.
Use the navigation keys to highlight and select the appropriate alarm menu selection. For
NIBP, the alarm menu selections are
NIBP Systolic Alarm, NIBP Diastolic Alarm,
and
NIBP MAP Alarm
.
240 mmHg
140 mmHg
120 mmHg
260 mmHg
150 mmHg
130 mmHg
Adult
Pediatric
Neonate
Limits
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...