![ZOLL Propaq MD Operator'S Manual Download Page 287](http://html2.mh-extra.com/html/zoll/propaq-md/propaq-md_operators-manual_3990877287.webp)
Defibrillator
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
A-3
* For all energy levels, accuracy is equal to either ±15% or 3 joules, whichever is greater.
The Propaq MD Rectilinear Biphasic waveform employs the same first and second phase
timing, similar first and second phase currents/voltages, and essentially the same mechanisms
for controlling defibrillation waveshape as the ZOLL R Series
®
defibrillator. The R Series and
Propaq MD defibrillation waveforms are considered substantially equivalent.
Figures A-1 through A-20 show the Rectilinear Biphasic waveforms that are produced when
the Propaq MD defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms
at each energy setting.
Table A-2.
Delivered Energy at Every Defibrillator Setting into a Range of Loads
Selected
Energy
Load
25
50
75
100
125
150
175
Accuracy*
1 J
1 J
1 J
1 J
1 J
1 J
1 J
1 J
±15%
2 J
1 J
2 J
2 J
2 J
2 J
2 J
2 J
±15%
3 J
2 J
3 J
3 J
3 J
3 J
3 J
3 J
±15%
4 J
3J
4 J
4 J
5 J
5 J
5 J
4 J
±15%
5 J
3 J
5 J
6 J
6 J
6 J
6 J
6 J
±15%
6 J
4 J
6 J
7 J
7 J
7 J
7 J
7 J
±15%
7 J
5 J
7 J
8 J
8 J
8 J
8 J
8 J
±15%
8 J
5 J
8 J
9 J
9 J
10 J
9 J
9 J
±15%
9 J
6 J
9 J
10 J
11 J
11 J
11 J
10 J
±15%
10 J
7 J
10 J
12 J
12 J
12 J
12 J
12 J
±15%
15 J
10 J
16 J
17 J
18 J
18 J
18 J
17 J
±15%
20 J
14 J
21 J
23 J
24 J
24 J
24 J
23 J
±15%
30 J
21 J
32 J
35 J
36 J
37 J
36 J
35 J
±15%
50 J
35 J
54 J
59 J
61 J
62 J
61 J
59 J
±15%
70 J
49 J
76 J
83 J
85 J
87 J
86 J
83 J
±15%
85 J
60 J
92 J
101 J
104 J
106 J
104 J
101 J
±15%
100 J
71 J
109 J
119 J
122 J
125 J
123 J
119 J
±15%
120 J
85 J
131 J
143 J
147 J
150 J
147 J
143 J
±15%
150 J
107 J
164 J
180 J
183 J
188 J
184 J
179 J
±15%
200 J
142 J
230 J
249 J
253 J
269 J
261 J
260 J
±15%
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...