11 M
ONITORING
I
NVASIVE
P
RESSURES
(IBP)
11-2
www.zoll.com
9650-0806-01 Rev. D
Warning!
•
If electrocautery is used, always avoid using any transducer with a conductive
(metal) case connected to its ground shield. Using a conductive transducer case that
is connected to its cable shield risks high-frequency burns at the ECG electrodes if
the transducer case becomes grounded to earth.
•
Normal alarm functions will detect complete disconnections of invasive pressure
transducers; however, the alarm functions will not detect a partial disconnection or
the use of some incompatible transducers. Use only approved transducers and ensure
that the transducers are connected properly.
•
Before you use the Propaq MD unit on a new patient, always turn it off for at least 2
minutes. This clears the previous patient’s trend values, alarm limit settings, and
NIBP cuff inflation pressure.
•
Use only ZOLL-approved IBP sensors. Use of non-approved sensors may result in
inaccurate IBP measurements.
IBP Setup
To take safe and accurate IBP measurements using the Propaq MD, you must perform the
following steps, each of which corresponds to a section in this chapter. Read each section
carefully before you perform IBP measurements.
1. Attach the invasive pressure transducer to the Propaq MD unit.
2.
Zero the transducer.
3.
Set the invasive pressure alarms (according to your organization’s standards).
4.
Select a label for the invasive pressure channel.
Attaching the Invasive Pressure Transducer
Follow these steps when attaching the invasive pressure transducer:
1. Inspect the transducer cable. If the cable shows any signs of wear, breakage, or fraying,
replace the cable. Replace the transducer dome, if necessary.
2.
Apply the transducer according to your organization’s procedures. Always refer the
manufacturer’s Directions for Use before using a transducer.
3.
If the transducer is a disposable unit with a separate cable, connect the transducer to the
transducer cable.
4.
Plug the transducer cable into one of the three six-pin IBP cable connectors on the side of the
Propaq MD unit.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...