Impedance Pneumography
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
A-15
Heart Rate Response to Irregular Rhythm
: (AAMI EC13-2002, section 4.1.2.1.e.)
Ventricular Bigeminy
: 80 BPM (expected)
Slow Alternating Ventricular Bigeminy
: 60 BPM (expected)
Rapid Alternating Ventricular Bigeminy
: 120 BPM (expected)
Bidirectional Systole
: 45 BPM (expected)
Tachycardia Response Time
:
Response time to tachycardia alarm is on average 3.4 seconds per AAMI EC-13-2002,
section 4.1.2.1.g, and IEC 60601-2-27, clause 6.8.2.bb.7. Response times include a 1.0
second display update interval.
Pacemaker Pulse Rejection:
(In accordance with AAMI EC13:2002, section 4.1.4 and IEC 60601-2-27:2005, subclause
50.103.13)
•
Pulses without overshoot: Rejects all pulses with amplitude of +2 mV to +700 mV and
duration of 0.1 ms to 2 ms, with no tail.
•
Pulses with overshoot: Rejects all pulses with amplitude of +2 mV to +700 mV and duration
of 0.1 ms to 2 ms, with overshoot up to 100 ms.
•
A-V sequential pulses: A-V sequential pacemaker pulses may not be rejected.
•
Fast ECG signals: Approximately 50% of ECG input pulses with a slew rate of 3 V/s RTI
may trigger the pacemaker pulse detector.
Electrosurgery Protection
: The Propaq MD is suitable for use in the presence of
electrosurgery as specified in IEC 60601-2-27. Burn hazard protection via a 1K current limiting
resistor contained in each ECG leadwire.
Impedance Pneumography
Displayed Data
: Numeric breath rate, Impedance waveform
Breath rate range
:
Adult, Ped: 2 to 150 breaths / minute
Neonates: 3 to 150 breaths / minute
Breath rate accuracy
: 2% or +/- 2 breaths / minute, whichever is greater
Displayed Breath Rate:
Average of last 10 breath-to-breath rates.
Leads
: Lead I (RA – LA), Lead II (RA – LL)
Sweep Speed
: 3.13, 6.25, 12.5 mm/sec
Alarm settings
: High, low and no breath rate alarm
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...