Configuring NIBP Alarms and Settings
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
8-13
By default, Smartcuf is
Enabled
.
There are situations where you may choose to disable Smartcuf. These may include
•
Very extreme motion artifact
•
Certain types of arrhythmias
•
Situations that prevent getting a ECG signal.
NIBP measurements can still be performed with Smartcuf disabled.
Start/Stop TurboCuf
Selecting
Start TurboCuf
starts Short-term Automatic (STAT) NIBP measurements. The
Propaq MD unit begins its first NIBP measurement, after which it continues to perform as
many NIBP measurements as possible over a 5-minute period.
Select
Stop TurboCuf
to immediately stop STAT measurements.
WARNING!
Repeated use of STAT measurements on the same patient over a short time interval can
affect blood pressure readings, limit circulation to the limb, and cause injury to the
patient.
Setting the NIBP Display Format
You can specify the appearance of the NIBP numeric display format. The display format allows
you to choose whether or not you will display MAP measurement (
M
) with the
Systolic (
S
) and Diastolic (
D
) measurements in one of the following formats:
•
S/D
•
S/D (M)
(default display)
•
(M) S/D
Note:
If a MAP alarm is active when the
S/D
format is selected, the display format will be
S/D (M)
until you clear the MAP alarm.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...