Synchronized Cardioversion Procedure
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
15–13
Unless otherwise configured, the unit automatically exits Sync mode after each shock. To
reactivate SYNC mode, press the
Sync
button on the front panel again. Changing the selected
energy levels does not cause the unit to leave SYNC mode.
Note:
The unit can be configured to stay in SYNC mode after defibrillation in
Defib/Pacer default settings in the Setup>Supervisor menu.
1 Select Energy Level
Press the
Select Energy
arrows up or down to select the desired energy level.These buttons are
located either on the front of the unit or on the STERNUM paddle.
Warning!
When used with pedi-padz, defibrillator energies must be set manually based on site-
specific institutional protocols for pediatric defibrillation.
2 Charge Defibrillator
Press the
CHARGE
button on the front panel or on the APEX paddle handle.
To increase or decrease the selected energy after you have pressed the
CHARGE
button, use
the defibrillator
Select
Energy
arrows on the front panel or sternum panel.
Caution
Changing the selected energy while the unit is charging or charged causes the defibrillator to
disarm itself. Press the
CHARGE
button again to charge the unit to the newly selected energy
level.
A charging message displays at the bottom of the screen, and a distinctive charging tone sounds
indicating that the unit is charging. The energy range bar graph on the right side of the display
highlights the charging progress until it reaches the selected energy. When the unit is fully
charged, the tone changes to a continuous charge ready tone, the highlighted energy bar graph
includes the selected energy, and the SHOCK button lights up.
or
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...