Page 54
MAINTENANCE
EQUIPMENT REQUIRED
Equipment Item
Function and Capabilities
1
Pressure Meter
(User supplied)
Monitors system pressure during a calibration procedure.
The Pressure Meter must be calibrated and have a minimum pressure range of
0-700 mmHg.
2
Pressure Source
(User supplied)
Supplies air pressure during a calibration procedure.
The pressure supply must provide an adjustably regulated minimum pressure
range of 0 to 700 mmHg and be no greater than the maximum monitored by
the pressure meter.
3
Calibration Hose Kit
(Provided)
Routes air pressure between a regulated pressure supply to a pressure meter
and the
A.T.S.
4000TS unit’s cuff ports.
The Calibration Hose Kit is supplied by Zimmer and included with the unit.
CAUTION:
The clamp on the calibration hose remains opened (not
clamped) during calibration.
CAUTION:
All calibration steps must be taken in the exact order to calibrate the unit. Failure to do so may
result in incorrect pressure readings while the unit is in operation.
3
2
1