
Page 38
INSTALLATION AND OPERATING INSTRUCTIONS
•
Touch the Stats button
to reveal and review statistics
of the previous pressure cycle:
■
The Pressure icon with right arrow FROM vertical bar
indicates the start pressure.
■
The Pressure icon with right arrow TO vertical bar indicates the ending pressure if the pressure was
adjusted during the procedure.
■
The LOP icon and text
LOP
/RTP
indicates the limb occlusion pressure (LOP) and recommended
Tourniquet pressure (RTP) that was determined if the LOP feature was used. All zeroes next to the LOP
text indicates that the LOP feature was not used.
■
The Time icon
indicates the total time the cuff was infl ated.
•
Touch the Okay button
to make no changes and return to the previous view. Touch the Reset button
to
reset the Stats to default. A confi rmation menu appears:
•
Touch Cancel button to make no changes and return to the Stats menu. Touch the Okay button
to
confi rm reset and return to the default GUI.