Page 44
INSTALLATION AND OPERATING INSTRUCTIONS
Warning
Message
Indicator
Light
Priority
Tone
Condition, Remarks and Appropriate Action
TIMED OUT
Constant
Yellow
Low
Condition: LOP determination timed out.
LOP determination is taking too long. Check LOP pulse sensor and retry. Replace LOP
pulse sensor with known good working sensor and retry. If problem persists, service the
unit.
LOP STOPPED
Constant
Yellow
Low
Condition: LOP determination prematurely stopped.
The LOP Cancel button was touched causing the LOP procedure to terminate. Press
the cancel button again to exit or Retry button to start LOP again. If problem persists,
service the unit.
HIGH LOP
Constant
Yellow
Low
Condition: High LOP.
Cuff pressure exceeded 300 mmHg during LOP measurement. Patient LOP is too high
for LOP measurement. Check LOP pulse sensor and retry. Replace LOP pulse sensor
with known good working sensor and retry. Do not use the LOP function, follow normal
Tourniquet procedures.
CAUTION:
Large surface area touches are detected, but ignored by the system.
ERROR CODES
SYSTEM FAILURE DETECTED, CALIBRATION ERROR DETECTED
and
BATTERY ERROR DETECTED
warning messages accompany one or more Error Codes. See
ALARM CONDITIONS
table for what the
warning messages mean.
Error Code
Warning
Message
Indicator
Light
Priority
Tone
Explanation / Appropriate Actions
E0001 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
EEPROM/ROM fail detected.
Specifi cation:
The EEPROM/ROM access was attempted or R/W error
occurred. The unit has shut down in a safe state.
Explanation/Action:
Microprocessor failed a ROM memory check. An
attempt was made to access or R/W to ROM (non-volatile memory)
and the values were lost. Cycle the ON/ STANDBY button. If problem
persists, note error code and contact manufacturer. If problem persists,
service the unit.
E0002
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Valve fail detected.
Specifi cation:
The unit has detected an improper valve combination or
missing valve/open valve. The unit has shut down in a safe state.
Explanation/Action:
A valve has failed during operation. A Safety
Monitor signal error was detected. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer. If problem
persists, service the unit.
E0003
XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit timing failure detected.
Specifi cations:
Internal system timing diagnostics feature has detected
an error. The unit has shut down in a safe state.
Explanation/Action:
Internal watchdog system detected a timing
malfunction. Cycle the ON/ STANDBY button. If problem persists, note
error code and contact manufacturer. If problem persists, service the unit.
E0004
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit failure detected.
Specifi cations:
Internal system diagnostics feature has detected a
hardware error. The unit has shut down in a safe state.
Explanation/Action:
A Safety Monitor hardware error was detected.
Cycle the ON/ STANDBY button. If problem persists, note error code and
contact manufacturer. If problem persists, service the unit.