Page 18
INSTALLATION AND OPERATING INSTRUCTIONS
•
The main TIME display should read “60” (the factory default set-point). The tile will automatically close in
approximately 5 seconds.
•
Within the 5 second time frame, touch the negative left-arrow button
-
to decrease or positive right-arrow
button
+
to increase the TIME set-point.
■
The set time can be maintained between 5 and 240 minutes in increments of 1 minute.
■
If an arrow button is continuously held, the increments will change by 5 minutes.
•
Touch the Okay button
to immediately remove the displayed tile or wait approximately 5 seconds and it will
automatically close.
•
Repeat
TIME set-point system test
for the blue Second Cuff TIME set-point.
CAUTION:
Anytime an asterisk (*) appears below the TIME value in the TIME tile, the data displayed is the
set-point. Set time will revert back to the default time setting when the unit is set to STANDBY.
3
Low Pressure Alarm Check
•
Connect a cuff and standard length hose set to the Main Cuff ports.
Touch the INFLATE (INFL) button to infl ate the cuff to the set-point of 250 mmHg: