Page 45
INSTALLATION AND OPERATING INSTRUCTIONS
Error Code
Warning
Message
Indicator
Light
Priority
Tone
Explanation / Appropriate Actions
E0006
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit pneumatic failure detected.
Specifi cations:
Internal system diagnostics feature has detected a
pneumatic error. The unit has shut down in a safe state.
Explanation/Action:
A Safety Monitor pneumatic error was detected.
Cycle the ON/ STANDBY button. If problem persists, note error code and
contact manufacturer. If problem persists, service the unit.
E0007
XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Valve fail detected.
Specifi cation:
The unit has detected an improper valve combination or
did not detect an expected valve combination. The unit has shut down in
a safe state.
Explanation/Action:
A Safety Monitor valve failure was detected. Cycle
the ON/ STANDBY button. If problem persists, note error code and
contact manufacturer. If problem persists, service the unit.
E0008
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit pneumatic enable failure
detected.
Specifi cations:
Internal system diagnostics feature has detected a
pneumatic enable failure. The unit has shut down in a safe state.
Explanation/Action:
A Safety Monitor pneumatic enable failure was
detected. Cycle the ON/ STANDBY button. If problem persists, note error
code and contact manufacturer. If problem persists, service the unit.
E0009
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Math fail detected.
Specifi cation:
The unit has detected a mathematical error within the
ALU. The unit has shut down in a safe state.
Explanation/Action:
Divide By Zero error. Result of math operation was
out of range. Cycle the ON/STANDBY Button. If problem persists, service
the unit.
E0010
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit failure detected.
Specifi cations:
Internal system diagnostics feature has detected an
error. The unit has shut down in a safe state.
Explanation/Action:
Unexpected software Interrupt occurred. Cycle
the ON/STANDBY button. If problem persists, note error code and
contact manufacturer. If problem persists, service the unit.
E0012
XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
EEPROM/ROM fail detected.
Specifi cation:
The EEPROM/ROM was attempted to be accessed and
an error was detected writing. The unit has shut down in a safe state.
Explanation/Action:
ROM check error detected a self-test. Cycle the
ON/STANDBY button. If problem persists, note error code and contact
manufacturer. If problem persists, service the unit.
E0014
BATTERY
ERROR
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Battery failed/error.
Specifi cation:
The unit has detected that the battery is missing,
disconnected or failed. The unit has shut down in a safe state.
Explanation/Action:
Verify a battery is connected and cycle the ON/
STANDBY button. If problem persists, note error code and contact
manufacturer.
E0015
BATTERY
ERROR
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Battery failed/error.
Specifi cation:
The unit has detected a non-Zimmer battery was installed
or the battery is no longer responding to commands. The unit has shut
down in a safe state.
Explanation/Action:
Verify a battery is connected and is a Zimmer
battery. Cycle the ON/STANDBY button. If problem persists, note error
code and contact manufacturer.