Page 22
INSTALLATION AND OPERATING INSTRUCTIONS
CAUTION:
•
If Advanced Leak Detection detected an excessive cuff leak during the infl ation period, the system will
indicate “LEAK DETECTED.”
•
Resetting the Stats also resets the cuff timer.
•
After approximately 5 seconds of inactivity the system will automatically close open menus and return
to its normal mode of operation.
SYSTEM FUNCTIONS
SETTINGS
Settings options include Defaults, Preferences, Cuff Test, and Service. The Service option is grayed out by default
but may be accessed by a qualifi ed biomedical engineer or other person thoroughly familiar with electronic medical
devices within the healthcare organization through the unlock sequence described in the
Maintenance
section.
To modify the system Settings, perform the following steps.
1
Touch the Settings button
and the Settings tool tray will slide up, revealing the Settings options:
2
If desired, touch the Cancel button to make no changes and return to the default GUI view.
CAUTION:
After approximately 5 seconds of inactivity the system will automatically close open menus and
return to its normal mode of operation
.
DEFAULTS
Defaults options include changing Pressure and Time settings. To change any Default setting, perform the following
steps.
To modify the default pressure or time settings for either cuff, follow the following steps.
1
Touch the Defaults menu button. The settings for Pressure and Time appear:
2
Touch the cuff-toggle button to switch between Main and Second Cuff.
3
Touch the negative left-arrow button - to decrease or positive right-arrow button + to increase the Default
Pressure or Time.