Page 23
INSTALLATION AND OPERATING INSTRUCTIONS
•
Default Pressure can be set between 50 and 400 mmHg in increments of 1 mmHg.
•
Default Time can be set between 1 and 240 minutes in increments of 1 minute.
•
If an arrow button is continuously held, the increments will change by 5 mmHg or minutes.
4
Touch the Okay button
to save the indicated Pressure and/or Time as the new default setting(s). The new
setting(s) will be the default every time the machine is turned on.
Touch the Cancel button to ignore any
changes and return to the Settings menu.
CAUTION:
•
No changes to defaults will be saved unless the Okay button
is touched to approve.
•
After approximately 5 seconds of inactivity the system will automatically close open menus and return
to its normal mode of operation
.
PREFERENCES
Preferences options include changing Alarm Volume, Display Brightness and/or Primary Language. To change any
Preference, perform the following steps.
1
Touch the Preferences menu button. The Preferences menu appears:
2
If desired, touch the Cancel button to make no changes and return to the default GUI view.
3
Select a Preference to change, as described below.
CAUTION:
After approximately 5 seconds of inactivity the system will automatically close open menus and
return to its normal mode of operation.
VOLUME SETTING
The Volume can be changed to accommodate different environmental conditions so that the audio alarm can be
reliably detected without being too intrusive in most situations. The
A.T.S.
4000TS has nine Volume levels ranging from
approximately 50 dB to 70 dB when measured at 1 meter from the center of the touch screen. To change the speaker
Volume, enter the Preferences menu and perform the following steps: