Page 47
INSTALLATION AND OPERATING INSTRUCTIONS
Error Code
Warning
Message
Indicator
Light
Priority
Tone
Explanation / Appropriate Actions
E0030 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Amplifi er pneumatic voltage failure detected.
Specifi cation:
The unit has detected a low voltage failure in the
pneumatic voltage supply. The unit has shut down in a safe state.
Explanation/Action:
Pneumatics voltage is out of range. Cycle the
ON/STANDBY button. If problem persists, note error code and contact
manufacturer.
E0031 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit failure detected an audio
failure.
Specifi cations:
Internal system diagnostics feature has detected a
speaker error. The unit has shut down in a safe state.
Explanation/Action:
Speaker not detected. Cycle the ON/STANDBY
button. If problem persists, note error code and contact manufacturer.
E0032 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
LOP signaling has failed.
Specifi cation:
The system is no longer detecting a stream of data from
the LOP circuitry. The unit has shut down in a safe state.
Explanation/Action:
Data no longer detected from the LOP circuit.
Cycle the ON/STANDBY button. If problem persists, note error code and
contact manufacturer.
E0033 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit failure detected an audio circuit
failure. The unit has shut down in a safe state.
Specifi cations:
Internal system diagnostics feature has detected an
error.
Explanation/Action:
Audio tone generation not detected. Cycle the
ON/STANDBY button. If problem persists, note error code and contact
manufacturer.
E0034 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Primary power supply rails out of specifi cation.
Specifi cation:
Loss of voltage on one of the DC power rails will cause a
failure to trigger. The unit has shut down in a safe state.
Explanation/Action:
Voltage lost from one or more of the DC power
rails. Power Supply (6V) failed. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer.
E0035 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Primary power supply rails out of specifi cation.
Specifi cation:
Loss of voltage on one of the DC power rails will cause a
failure to trigger. The unit has shut down in a safe state.
Explanation/Action:
Voltage lost from one or more of the DC power
rails. Power Supply (3.3V) failed. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer.
E0036 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Primary power supply rails out of specifi cation.
Specifi cation:
Loss of voltage on one of the DC power rails will cause a
failure to trigger. The unit has shut down in a safe state.
Explanation/Action:
Voltage lost from one or more of the DC power
rails. Power Supply (1.8V) failed. Cycle the ON/STANDBY button. If
problem persists, note error code and contact manufacturer.
E0037 XXXX
1
SYSTEM
FAILURE
DETECTED
Constant Red
Technical
Failure
Constant tone
Condition:
Watch dog timer safety circuit failure detected.
Specifi cations:
Internal system diagnostics feature has detected a clock
error. The unit has shut down in a safe state.
Explanation/Action:
Internal watchdog system detected a malfunction.
Safety Circuit Clock not functioning properly. Cycle the ON/STANDBY
button. If problem persists, note error code and contact manufacturer.