CARE AND SERVICE
ZEISS
Care
Colibri 7
32
423052-7344-001
02/2017
5
CARE AND SERVICE
5.1
Care
Disconnect the power plug of the desktop power supply unit prior to cleaning!
Clean labels and desktop power supply unit only dry.
•
Clean lamp module, control panel and LED Trigger Box only with a slightly moist cloth (not dripping
wet). Do not allow moisture to penetrate the inside of the devices.
•
All labels on the components and the desktop power supply unit may only be cleaned using a dry
cotton cloth. Do not allow moisture to penetrate the inside of the desktop power supply unit.
When using the device in warm and humid climatic zones, pay attention to the following instructions:
•
Store the device in bright, dry and well-ventilated rooms with a humidity of less than 65 %; store
particularly sensitive optical components and accessories, such as objectives and eyepieces, in a dry
cabinet.
•
If the microscope is to be stored in closed containers for a prolonged period, fungicide-soaked cloths
should be placed in the containers to prevent mould.
The risk of growth of fungus on opto-mechanical instruments always exists in the following
conditions:
−
relative humidity > 75 % and temperatures b15 °C and +35 °C for more than
three days
−
installation in poorly-ventilated, dark rooms and
−
dust deposits and fingerprints on optical surfaces.