OPERATION
Carl Zeiss
Mounting the Filter Wheel
Cell Observer SD
58
423638-7044-001
M60-2-0021 e 01/2009
4.8
Mounting the Filter Wheel
•
Use the key switch (Fig. 4-1/
2
) to turn the
spinning disc unit off.
•
Remove the power plug from the power pack.
•
Release the clamping screw (2.5" Allen key)
(Fig. 4-11/
1
) at the spinning disc unit.
•
Extract the filter insert (Fig. 4-11/
2
) from the
front.
•
Introduce the filter wheel (Fig. 4-12/
2
) into the
spinning disc unit until mechanical stop po-
sition.
•
Tighten the clamping screw (2,5" Allen key)
(Fig. 4-12/
1
).
•
Connect the filter wheel to the CSU-X1 control
box (refer to section 3.6.2).
Exchange of filter(s) in filter wheel:
Necessary actions for replacement of a filter or filters
are described in the "Filter Wheel CSUX1FW"
manual (supplied) (IM
85A07C27-01E) from
Yokogawa. However, users are advised to contact
the Service Department of Carl
Zeiss
MicroImaging GmbH or specially
authorized
expert
personnel for necessary filter replacement.
Fig. 4-11
Removal of emission filter insert
Fig. 4-12
Mounting the filter wheel