START-UP
Cell Observer SD Coupling a Customer-Provided Laser to the Cell Observer SD System
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
47
3.9
Coupling a Customer-Provided Laser to the Cell Observer SD System
Customers using their own lasers will be solely responsible for overall laser safety of the Cell
Observer SD system. They shall also check for further required laser safety precautions and
validity of all laser warning labels affixed to the system, including the microscope.
The Manufacturer will refuse liability for modifications performed on the Cell Observer SD
system.
To operate the system with a total laser power > 300 mW, further laser safety precautions will
be required for the microscope.
Where lasers with a total laser power > 300 mW or other wavelengths are used, the Cell
Observer SD protection system must be reviewed for protective effect.
Notify your Laser Safety Officer before proceeding with the initial system start-up.
Valid national laser protection requirements must be met.
Note:
This work step is only required for integration of a Customer laser. A special optical fiber cable is
necessary for coupling a laser to the Cell Observer SD system (refer to section 2.6 on page 31). If you
intend to mount your own laser, you should consult the related manufacturer manual for details and
instructions regarding installation and alignment of the coupling device in front of the laser.
For operation with your own laser, the laser shutter for the laser module must always be
mounted in a position between the coupling device and the laser outlet. The laser beam must
be completely blocked by the shutter flag when in lower (de-energized) position.
Switch the laser off before mounting the optical fiber!