START-UP
Carl Zeiss
Safety Check before Laser Start-up
Cell Observer SD
48
423638-7044-001
M60-2-0021 e 01/2009
3.10
Safety Check before Laser Start-up
Check for proper function of the laser safety device before you proceed with initial start-up
actions or after each re-installation of the Cell Observer SD system.
1.
Activate the laser safety device by turning the microscope on.
2.
Remove the tube as described in the Axio Observer.Z1 Microscope Operating Manual (B 46-111) to
be able to see the eyepiece shutter.
3.
Check the following action items now:
Action System
state
Turn reflector turret until it is unlocked or remove the
turret from the turret compartment
•
Laser shutter closed
•
Eyepiece shutter open
Reflector turret locked inside turret compartment and
rotate turret into locked position
•
Laser shutter open
•
Eyepiece shutter closed
Remote connector detached from distribution box for laser
safety signals
•
Laser shutter closed
•
Eyepiece shutter open
Remote connector plugged in distribution box for laser
safety signals
•
Laser shutter open
•
Eyepiece shutter closed
Transmitted-light arm of stand in beam position
•
Laser shutter closed
•
Eyepiece shutter open
Transmitted-light arm of stand in off-beam position
•
Lasers shutter open
•
Eyepiece shutter closed
Distribution box for laser safety signals not connected to
microscope
•
Laser shutter closed
•
Eyepiece shutter open
Some plug-in positions of distribution box for laser safety
signals unused
•
Laser shutter closed
•
Eyepiece shutter open
If your safety device is not found to function as described above, shut the system down, mark it
clearly as defective and contact the Service Department of Carl Zeiss MicroImaging GmbH
immediately.
Attach the microscope tube again after successful safety check.