INSTRUMENT
DESCRIPTION
Carl Zeiss
Instrument Description and Main Features
Cell Observer SD
28
423638-7044-001
M60-2-0021 e 01/2009
2.3.6
Basic Laser Safety Kit
The basic laser safety kit includes an eyepiece shutter (factory-installed or retrofitted by service personnel),
a laser safety board for the stand, a distribution box and sticker labels.
The distribution box for laser safety (Fig. 2-8/
3
) must be connected to the rear of the stand (Fig. 3-9)
using a 15-pole sub-D connector (Fig. 2-8/
1
).
The four jack terminals of this distribution box (Fig. 2-8/
2
) are available for connection to the laser
protection cover or laser protection incubator, the laser shutter and the multi-laser module.
Jumpers must be installed onto unused jack terminals of the distribution box. This is necessary to enable
release motion for the laser shutter inside the laser module.
In addition to that, the distribution box contains a connection for a door interlock switch (Fig. 2-8/
4
) This
connection is provided with a factory-installed crossover connector.
Fig. 2-8
Distribution box for laser safety