CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Carl Zeiss
Replacement of Fuses
Cell Observer SD
62
423638-7044-001
M60-2-0021 e 01/2009
5.4
Replacement of Fuses
Disconnect the power plug in any case before replacing the fuses.
Axio Observer, HBO, XBO, CTI-Controller, Tempcontrol, Definite Focus: consult pertaining manual.
CSU-X1 spinning disc unit: There are no fuses with external access for replacement. Please contact the
Service Department of Carl Zeiss MicroImaging GmbH.
Signal distribution box:
The fuse compartment is combined with the inlet
connector. It contains two fuses.
•
Use main power switch to turn the module off.
•
Pull the main power plug.
•
2
) out from the front.
Use a small screwdriver for convenience.
•
Extract fuses from the fuse holder and replace
them with new fuses of type T 1A/H/250 V.
•
Insert fuse holder into the fuse compartment
until mechanical stop position (Fig. 5-1/
1
).
•
Re-connect main power plug.
Multi laser module:
The fuse compartment is combined with the inlet
connector and the main power plug. It contains
two fuses.
•
Use main power switch to turn the module off.
•
Pull the main power plug.
•
2
) out from the front.
Use a small screwdriver for convenience.
•
Extract fuses from the fuse holder and replace
them with new fuses of following types:
T
3.15A/E/250
V for 230
V operation and
T 6.3A/E/250 V for 120 V operation.
•
Insert fuse holder into the fuse compartment
(Fig. 5-2/
1
) until mechanical stop position.
•
Re-connect main power plug.
Fig. 5-1
Replacement of fuses in signal
distribution box
Fig. 5-2
Replacement of fuses in multi laser
module