![background image](http://html1.mh-extra.com/html/zeiss/cell-observer-sd/cell-observer-sd_operating-manual_3389359033.webp)
INSTRUMENT
DESCRIPTION
Cell Observer SD
Technical Data
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
33
Operating data
Axio Observer.Z1, motorized with external VP 231 power supply:
Intended site..........................................................................................................................closed rooms
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Electrical safety .............................................................in compliance with DIN EN 61010-1 (IEC 61010-1)
under consideration of CSA and UL directives
Safety of laser devices...................................................in accordance with DIN EN 60825-1 (IEC 60825-1)
Overvoltage category............................................................................................................................... II
Radio interference suppression ..........................................................in accordance with EN 55011 class B
Noise immunity................................................................................... in accordance with DIN EN 61326-1
Line voltage ranges of Axio Observer.Z1
External VP 231 power supply, stabilized................................... 100 to 127, 200 to 240 VAC (± 10 %)
Change of instrument voltage setting is not necessary!
Line frequency ........................................................................................................................50/60 Hz
Power consumption with internal power supply units:
Axio Observer.Z1 with external VP231 power supply unit ......................................................... .280 VA
Power supply unit for XBO 75, 100 - 240 V AC ......................................................................... 130 VA
Power supply unit for HBO 100, 100 - 240 V AC....................................................................... 155 VA
Fuses according to IEC 127
Axio Observer.Z1 stand
100 ... 240 V AC...................................................................................... T 6.3 A/H; 250 V; 5 × 20 mm
Light sources
Halogen lamp .................................................................................................................. 12 V / 100 W
Controllability of light source ............................................................ continuously, approx. 3 V to 12 V
Mercury-vapor short-arc lamp HBO 103 W/2.............................................................................. 100 W
Power ........................................................................................................................................ 100 W
Optical-mechanical data
Stand with stage focusing drives - coarse focusing drive..................................................2 mm/rotation
and fine focusing drive.................................................................................................0.2 mm/rotation
Fine scale ................................................................................................................. 1 µm/scale division
Total focusing range ..................................................................................................... approx. 10 mm
Exchange of objectives ................................................................................................ via 6x lens turret
Objectives ................................................................................................ with M27x0.75 screw mount
Eyepieces, tube diameter............................................................................................................ 30 mm
and with field-of-view number ..........................................................................................................23
Summary of Contents for Cell Observer SD
Page 1: ...l j l pa...