background image

DO NOT save each section as a 

new file.  Use the workflow to add 

new sections, as described above, 

until all the sections in an animal 

are traced and counted. 

NOTE: 

One animal = one file.

and past the bottom of the tissue until 

it is completely out of focus and then 

slowly focus back up until something 

just comes into focus.  This is the bottom 

of the tissue.
Note: If more than one person is doing 

the counting for one experiment, all the 

counters should sit down and come to 

a consensus as to what is the top and 

bottom of the tissue because if they 

choose differently, the difference will be 

reflected in the results.
As far as choosing cells to count (place 

a marker on), pick a point of the cell 

(such as the top of the cell, the widest 

perimeter of the cell, the nucleus, the 

widest perimeter of the nucleus or the 

nucleolus) and when that point of the 

cell comes into focus, place a marker on 

it, as long as it falls into the rules of the 

counting frame discussed in step 6 and 

falls within the dissector height (shown 

as green in the Z meter).  If this point of a 

cell comes into focus while in the guard 

zone (shown as red in the Z meter) the 

software will block counting it.   
Once all the sites have been visited, click 

on Add New Section in the workflow 

below the blue start counting arrow.  

The software will then return you to step 

2 in the workflow for the next section.  

Follow the workflow down to Step 11 

again.  Continue counting each sec-

tion and adding new sections until all 

the sections the particular animal are 

completed.

Step 12: View Sampling Results
Once counting is complete, the best way 

to view the results is to click on Display 

Probe Run List.  Once this list pops up, 

highlight all the like contours in each 

section.  For example, if there is only one 

contour per section, highlight all of them 

(hold down the Ctrl key to select more 

than one) and click on View Results. If 

the sections have  mirrored contours  

Left and Right, highlight all the  left con-

tours and View Results.  Later, highlight 

all the right contours and View Results 

for that.   To see the results for the com-

bined contours, highlight all the right 

and left contours and View Results. 

Interpretation of Results
The first items to appear in the results 

window are the parameters for your 

study.  Marker will show the estimated 

total cell numbers.  Do not be alarmed 

if some of the results are 0.  Here is the 

explanation of the different results:

Estimated total by optical fractionator:

  

If you manually entered the thickness 

at the beginning of counting (and did 

not measure while counting), then you 

use the first output in display probe run 

list. This calculates the cell number and 

bases the counts on the manual thick-

ness that you entered prior to doing the 

cell counts.  This is the *LEAST accurate 

estimated value* unless you are 100% 

sure of the thickness of all of your sec-

tions.  If you did not enter the thickness 

manually this number will equal zero in 

the results, which can be changed by 

clicking on Edit Mounted Thickness. 
  

Estimated total mean measured thick-

ness:

  This second value calculates cell 

number based on the average thick-

ness of the tissue.  You would use this 

if you measured section thickness at 

every sampling site (or periodically dur-

ing counting) and it will calculate the 

counts based on the mean thickness 

throughout the tissue. This is the *MOST 

accurate* of the cell estimates and we 

recommend that our customers mea-

sure at every site and use this estimate 

for their data analysis. 
 

Estimated total number by weighted 

thickness:

  This third value calculates 

cell number by taking into account very 

wavy tissue.  So if you noticed that there 

was a lot of variation between sampling 

sites in terms of section thickness, you 

should use this cell count estimate.

You can also see the various CEs by 

clicking on them.  Clicking on Planim-

etry will display the volume and area for 

the structure.

Software Bugs: (Save Often)

• If No Live Image

If no live image appears check camera activ-

ity LED on top of camera body; if it is green 

it is on but not talking to SI or NL.  If Green 

do this: (Amber flicker = Active)
1. Close software
2. Turn off Master switch
3. Restart Computer

• If Software locks up

1. Turn off master switch
2. Restart Computer
3. Turn on Master switch
4. Login to clear logbooks; depending on 

the crash, ITG logbooks may not close until 

a user has logged in to the computer.

• Before You Leave:

Close the software and restart the computer 

instead of log out. This will ensure all back-

ground processes are stopped, improving 

stability for users after you.

Summary of Contents for AxioImager A1

Page 1: ...Instructions for Zeiss AxioImager A1 Microscope for Stereology Basic operation instructions as of 10 1 2009 The Stereology Workstation By Jon Ekman...

Page 2: ...fluorescence viewing Press the RL Key Set Reflector Turret to 2 Eyes DUAL 470EX 560EX 515EM 620EM Adjust the Incident Iris Stop Slider until light intensity is adequate Find Focus preferably on a dye...

Page 3: ...crews 3 5 Open luminous field diaphragm 1 to edge of field 6 Adjust contrast using condenser aperture diaphragm slider 4 In most cases adjust aperture diaphragm slider to mark matching the NA of the o...

Page 4: ...oscope and in the software Align for Koehler Illumination Grid Tune Calibration 8 Select Tools Grid Tune Current Lens 9 Tell the software the box size Large grid 250mm Small grid 25mm 10 Drop the anch...

Page 5: ...o ing a pilot study will ensure that quality data is generated by obtaining the best estimates of total cell number Step 1 Set up the subject 1 Number of sections to count De fault is 1 new sections c...

Page 6: ...eing counted NOTE Section Thickness measurements Guard Zones Step 8 Define Dissector Options Set Guard zones so that an over or under estimation of cell totals does not occur Do not count cells that f...

Page 7: ...of Results The first items to appear in the results window are the parameters for your study Marker will show the estimated total cell numbers Do not be alarmed if some of the results are 0 Here is t...

Page 8: ...with setting exposure Gain and other camera specific options 6 Adjust Exposure and white balance using More button in the Camera Set tings window another window will pop up with the basic Olympus Pic...

Page 9: ...3 Plan Apochromat excellent quality cor rect for spherical aberration and corrected for chromatic aberration in red blue and green wavelengths 4 EC Epiplan Enhanced Contrast Univer sal objectives for...

Page 10: ...as the Z drive in the windows environment ITG is equipped with gigabit ethernet and saving to the Z drive is relatively fast Also this permits easy access to the data from any computer in the world u...

Page 11: ...hers 1998 Michalet X Kapanidis A N Laurence T Pinaud F Doose S Pflughoefft M and Weiss S The power and prospects of fluorescence microscopies and spectrosco pies Annu Rev Biophys Biomolec Struct 32 16...

Reviews: