background image

Stereology with the Optical Fractionator Workflow

Tissue Preparation
The optical fractionator requires fairly 

thick tissue.  Thicker tissue allows one to 

distinguish between cells that are on top 

of each other while focusing through 

the tissue making it easier to see the 

cells clearly as they start to come into 

focus. Optimal thickness for this method 

is 20 microns after accounting for shrink-

age from tissue processing.
Options here are:
 1.  Cut thicker sections.
 2.  Dry the tissue as little as possible.
 3.  Avoid unnecessary alcohol steps. 
Collect serial tissue sections without 

losing any sections along the way.  The 

Optical Fractionator requires a series of 

sections, such as every 3rd section.  The 

interval will depend on the size of the 

structure and how many cells the struc-

ture contains; this can be determined 

with a pilot study.

Do a Test Run or Pilot Study
A pilot study will help decipher the best 

parameters to use for each experiment 

including counting frame size, grid size, 

and serial section interval. Due to vari-

ability across animals, a pilot study may 

be required on one animal from each 

group within the experiment. 
With the Optical Fractionator the target 

is to count no less than 200 cells per 

animal. This can be achieved by count-

ing 10 sections with 10 sampling sites 

per section and on average 2 cells per 

sampling site. Make sure the Coefficient 

of Error (CE) is below 0.1.  The best way 

to approach a pilot study is to err on 

the side of over-counting.  Once cells 

counts have been completed and the 

CE has been calculated for the test 

animal, you can decide whether or not 

to adjust the parameters. 
Once the results of a pilot study, have 

been reviewed, adjustments to the pa-

rameters can be made in order to count 

more cells or fewer cells as necessary.  
Ways to DECREASE the number of 

cells:
 1.  Make the counting frame smaller.
 2.  Decrease the grid size creating fewer 

sampling sites.
 3.  Increase the section interval, for 

example, counting every 4th section 

instead of every 3rd. 
Ways to INCREASE the number of 

cells you count:
 1.  Make the counting frame bigger.
 2.  Make the grid size smaller so that 

you visit more sites.
 3.  Decrease the section interval, for 

example, counting every 3rd section 

instead of every 4th.   
Once good parameters for an experi-

ment have been determined, they can 

be saved and quickly pulled up in the 

workflow for subsequent animals.  Do-

ing a pilot study will ensure that quality 

data is generated by obtaining the best 

estimates of total cell number. 
Step 1: Set up the subject
 1.  Number of sections to count:  De-

fault is 1;  new sections can be added as 

you go.

 2.  Section’s cut thickness:  How thick the 

tissue was cut at.
 3.  Section Evaluation Interval:  This is 

the interval between the sections to 

count.  Every other section =2, every 3rd 

=3 ...   
 4.  Starting Section Number:  Default; 

start counting at section 1.  The first 

section is defined as the first section that 

displays the desired morphology.
 5.  Z-value of First Section: This will 

always be 0.00 at section 1.  If starting on 

a section other than one, it will be an in-

terval of the cut thickness.  For example 

if tissue was cut at 50 microns and you 

are starting on section one, the Z value 

=0.00, if starting on section 2 the Z value 

=50, if starting on section 3, the Z value 

= 100, and so on.

Step 2: Set microscope to Low Magni-

fication
Choose a low power objective (2.5x, 5x 

or 10x) best fitted for tracing a contour 

around your region of interest (ROI).  

Step 3:  Trace your Region of Interest 

(ROI)
Select a contour to use and then trace.  

Use this same contour choice for the 

same region of interest in subsequent 

sections.  More than one contour can 

be traced to count different regions in a 

structure or different structures.  Con-

tours can be named uniquely.
Hint: Name contours after your area of 

interest so they can quickly be selected 

in the Display menu. Click on Display Set-

tings and then the contour tab.  Click on 

the contour to rename.  The change will 

be reflected in the workflow.
Step 3 is a good place to edit a contour.  

Go to the Edit menu and click on Select 

Objects.  Then click on the contour and 

it will be highlighted with white squares.  

Once it is highlighted, right click for 

various editing tools. Remember to right 

click and Exit the Selection Tool once 

editing is complete.

Step 4: Set Microscope to High Magni-

fication
Select a high power objective that will 

be good for counting (40x-oil, 63x-oil or 

These Parameters should be set so 

that you end up counting at least 

200 cells and/or a CE of less than 0.1 

is achieved without spending time 

counting more cells than you need to. 

Also, do not change parameters for 

every section within an animal!  All 

parameters must be kept constant 

throughout all the sections of the 

animal for the calculations to be 

valid.  It is best to keep the same pa-

rameters for each group of animals 

as well. 

NOTE: 

counting frame size, grid size, and serial section interval

Information below is derived from MBF Bioscience correspondence

Summary of Contents for AxioImager A1

Page 1: ...Instructions for Zeiss AxioImager A1 Microscope for Stereology Basic operation instructions as of 10 1 2009 The Stereology Workstation By Jon Ekman...

Page 2: ...fluorescence viewing Press the RL Key Set Reflector Turret to 2 Eyes DUAL 470EX 560EX 515EM 620EM Adjust the Incident Iris Stop Slider until light intensity is adequate Find Focus preferably on a dye...

Page 3: ...crews 3 5 Open luminous field diaphragm 1 to edge of field 6 Adjust contrast using condenser aperture diaphragm slider 4 In most cases adjust aperture diaphragm slider to mark matching the NA of the o...

Page 4: ...oscope and in the software Align for Koehler Illumination Grid Tune Calibration 8 Select Tools Grid Tune Current Lens 9 Tell the software the box size Large grid 250mm Small grid 25mm 10 Drop the anch...

Page 5: ...o ing a pilot study will ensure that quality data is generated by obtaining the best estimates of total cell number Step 1 Set up the subject 1 Number of sections to count De fault is 1 new sections c...

Page 6: ...eing counted NOTE Section Thickness measurements Guard Zones Step 8 Define Dissector Options Set Guard zones so that an over or under estimation of cell totals does not occur Do not count cells that f...

Page 7: ...of Results The first items to appear in the results window are the parameters for your study Marker will show the estimated total cell numbers Do not be alarmed if some of the results are 0 Here is t...

Page 8: ...with setting exposure Gain and other camera specific options 6 Adjust Exposure and white balance using More button in the Camera Set tings window another window will pop up with the basic Olympus Pic...

Page 9: ...3 Plan Apochromat excellent quality cor rect for spherical aberration and corrected for chromatic aberration in red blue and green wavelengths 4 EC Epiplan Enhanced Contrast Univer sal objectives for...

Page 10: ...as the Z drive in the windows environment ITG is equipped with gigabit ethernet and saving to the Z drive is relatively fast Also this permits easy access to the data from any computer in the world u...

Page 11: ...hers 1998 Michalet X Kapanidis A N Laurence T Pinaud F Doose S Pflughoefft M and Weiss S The power and prospects of fluorescence microscopies and spectrosco pies Annu Rev Biophys Biomolec Struct 32 16...

Reviews: