background image

Optical Fractionator Workflow Continued...

100x-oil).  Objective choice will depend 

on tissue type and staining. Note: objec-

tives with magnification lower than 40x 

will not give you enough depth of view 

to count accurately especially when cells 

are close to each other in Z.

Step 5: Measure Section Thickness
The calculations for Optical Fractionator 

require the measured thickness of the 

tissue post processing.  Ideally, measure 

the tissue thickness at every sampling 

site for the most accurate cell estimates.  

Measured thickness from all counting 

sites will then be averaged and used in 

the calculations. 
 1.  Check the Measure the section while 

counting option and leave the other op-

tions unchecked. 
 2.  Set your Evaluation Interval for 

measuring to 1.  Every site will require a 

thickness measurement.

Step 6: Define the Counting Frame 

Size
Set counting frame size to have 1-5 

countable objects within.
Counting Frame tip:  Choose a unique 

identifying point of your cell or object 

of interest, such as cell top, nucleus 

top, nucleolus, widest cell perimeter, or 

widest nucleus perimeter.  Consistently 

count objects only when the identifying 

point is in focus.
Remember, also that the counting frame 

can be placed during this step and the 

software will always display it where you 

placed it. Click and drag in the center, 

or anywhere on the screen and that will 

be the location (click) and size (drag) 

that will be used while using the optical 

fractionator.

Step 7: Define SRS Grid Layout
Set the grid size so that there are ap-

proximately 10 counting frames in the 

region of interest (ROI).  Remember, this 

is just an approximation.  The number 

of counting frames per ROI will never 

be constant, because contours change 

size in each section and because the SRS 

Grid Layout is randomly thrown down 

before every run to help reduce Bias. 

If the sections have  a history of hav-

ing very uniform thickness, choose to 

set the Evaluation Interval to be larger 

than one.  Possible options could be  

at every 3rd sampling site.  This can 

be done to save time.
As for Guard Zones, it is recommend-

ed that the guard zone be a minimum 

of half the thickness of the object  

being counted.

NOTE: 

Section Thickness measurements & Guard Zones

Step 8: Define Dissector Options
Set Guard zones so that an over- or 

under-estimation of cell totals does not 

occur.  Do not count cells that fall into 

the guard zone.  Guard zones are used 

to ensure objects counted are not from 

an area where potential knife damage 

can occur either by being cut in half, or 

plucked from the tissue during micro-

tomy.  In practice, we recommend that 

the guard zone be a minimum of half the 

thickness of the object you re count-

ing.  The dissector height will define 

the depth in which you will count cells.  

This will also depend on how thick the 

tissue is.  For example, a section that is 

20 microns thick after processing, could 

have a top guard zone of 4 microns, with 

a dissector height of 12 microns, leav-

ing by default a bottom guard zone of 4 

microns.  If tissue varies a lot in thickness, 

choose a dissector and guard zones that 

will fit into the thinnest piece of tissue.

Step 9: Save Sampling Parameters
Save the sampling parameters that were 

determined in the pilot study.  Then for 

subsequent animals, choose the saved 

sampling parameters from Step 1: Set up 

the Subject and skip steps 5-9.

Step 10: Select Markers for Counting
Choose a marker for counting and use 

this same marker for subsequent sec-

tions.  Change the name of the marker 

to match the type of cells counted by 

going to the Display menu, Display Set-

tings, and the Marker Tab.  Click on the 

marker and rename it.  This name will 

then appear in the marker tool bar and 

will be easily visible for you to select it 

for counting. On reload of a Data set the 

software will ask to use marker names if 

they are not present.

Step 11: Count Objects
In this step  choose from a list of the 

sections and contours that were drawn 

for each section, by highlighting the con-

tour of a specific section to count and 

click on the blue arrow.  The stage will 

move to the first site and the software 

will prompt  to measure the top and the 

bottom of the tissue.  To find the top of 

the tissue focus up past the top of the 

tissue until it is completely out of focus.  

Slowly focus back down onto the tissue 

and as soon as something on the tissue, 

whether it is a cell or connective tissue 

comes into focus. This is the top of the 

section.  To measure the bottom of the 

section, focus all the way down through 

Summary of Contents for AxioImager A1

Page 1: ...Instructions for Zeiss AxioImager A1 Microscope for Stereology Basic operation instructions as of 10 1 2009 The Stereology Workstation By Jon Ekman...

Page 2: ...fluorescence viewing Press the RL Key Set Reflector Turret to 2 Eyes DUAL 470EX 560EX 515EM 620EM Adjust the Incident Iris Stop Slider until light intensity is adequate Find Focus preferably on a dye...

Page 3: ...crews 3 5 Open luminous field diaphragm 1 to edge of field 6 Adjust contrast using condenser aperture diaphragm slider 4 In most cases adjust aperture diaphragm slider to mark matching the NA of the o...

Page 4: ...oscope and in the software Align for Koehler Illumination Grid Tune Calibration 8 Select Tools Grid Tune Current Lens 9 Tell the software the box size Large grid 250mm Small grid 25mm 10 Drop the anch...

Page 5: ...o ing a pilot study will ensure that quality data is generated by obtaining the best estimates of total cell number Step 1 Set up the subject 1 Number of sections to count De fault is 1 new sections c...

Page 6: ...eing counted NOTE Section Thickness measurements Guard Zones Step 8 Define Dissector Options Set Guard zones so that an over or under estimation of cell totals does not occur Do not count cells that f...

Page 7: ...of Results The first items to appear in the results window are the parameters for your study Marker will show the estimated total cell numbers Do not be alarmed if some of the results are 0 Here is t...

Page 8: ...with setting exposure Gain and other camera specific options 6 Adjust Exposure and white balance using More button in the Camera Set tings window another window will pop up with the basic Olympus Pic...

Page 9: ...3 Plan Apochromat excellent quality cor rect for spherical aberration and corrected for chromatic aberration in red blue and green wavelengths 4 EC Epiplan Enhanced Contrast Univer sal objectives for...

Page 10: ...as the Z drive in the windows environment ITG is equipped with gigabit ethernet and saving to the Z drive is relatively fast Also this permits easy access to the data from any computer in the world u...

Page 11: ...hers 1998 Michalet X Kapanidis A N Laurence T Pinaud F Doose S Pflughoefft M and Weiss S The power and prospects of fluorescence microscopies and spectrosco pies Annu Rev Biophys Biomolec Struct 32 16...

Reviews: