7 Maintenance
Instruction for use
AT 030
7.4 Checking the device
44 / 60
319300-9120-000-GA-GB-240320
ð
Then, tighten the left set screw.
ð
The left set screw is only accessible if the tonometer holder
is pivoted approx. 30° to the right from the stop notch.
u
If the measuring prism is to the left of the center of the field of
view, proceed as follows:
ð
Loosen the left set screw slightly.
ð
Then, tighten the right set screw. The left set screw is only
accessible if the tonometer holder is pivoted approx. 30° to
the right from the stop notch.
u
Adjust alternately until the measuring body can be seen in the
center of the visual field.
u
Then retighten both set screws alternately.
Checking the achieved adjustment status
u
Perform a final check of the adjustment status.
u
If necessary, repeat the steps in sections Setting the correct
height in the field of view (height adjustment A) and Setting
the lateral middle position in the field of view (lateral
adjustment D).
u
If required, move the tonometer to the left alignment hole of
the base plate and check the adjustment of the left eyepiece.
7.4 Checking the device
After transport of the device, check the condition of the device as
follows:
Checking while mounted on slit lamp
Action
u
Move the tonometer to the measuring position.
u
Set the scale reading to 0. The measuring prism can easily be
moved back and forth.
u
Set the scale reading to 1. The measuring arm with the
measuring prism must move immediately towards the front end
position.
NOTE! If the measuring arm with the measuring prism does
not move to the front end position, discontinue using the
device. Consult ZEISS Service.
Please also observe the infor-
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...