Instruction for use
8 Troubleshooting
AT 030
319300-9120-000-GA-GB-240320
47 / 60
8 Troubleshooting
Action
1. The axis of the measuring prism must coincide with the normal
to the corneal surface. This is the case if the fluorescein
semicircles are symmetrical to the separating line.
The two fluorescein semicircles are visible in the center of the
eyepiece and touch each other at their inner edges.
Figure 14: Correct measuring pattern
2. If this condition is not met, patterns as shown in the following
figure will appear. These patterns require the slit lamp and thus
the measuring prism to be repositioned laterally and vertically.
Figure 15: Measuring patterns requiring correction of the tonometer
position
Possible faults and their remedies are described below.
Fault
Eyepiece
image
Remedy
Fluorescein ring too wide/too small
Fluorescein ring too wide:
The measuring prism was not dried
after cleaning or the eyelids
touched the measuring prism
during measurement. In this case,
the reading is higher than the
actual pressure.
Pull the slit lamp
back and dry the
measuring prism
with a lint-free
cloth (e.g. made
from cellulose)
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...