Instruction for use
7 Maintenance
AT 030
7.3 Adjusting the swivel-mounted tonometer holder with round mounting
319300-9120-000-GA-GB-240320
41 / 60
Setting the correct height in the field of view (height
adjustment A)
u
Set the magnification of the slit lamp microscope to a high
magnification (e.g. 20x or 25x).
u
Loosen the two clamping screws by approx. ½ turn.
u
Unscrew the adjusting screw at the upper end of the rod by
approx. 5 mm.
u
Press on this adjusting screw and slide the rod with tonometer
downwards, until the base plate is resting on the pivot joint of
the slit lamp.
u
Move the tonometer with measuring prism into the vertical
center of the field of view by turning the adjusting screw to the
right. Hold the tonometer, so that it cannot rotate laterally.
Ensure the correct alignment of the measuring prism.
u
Firmly retighten the two clamping screws alternately.
u
Retighten the adjusting screw carefully.
Adjust height
Height set correctly
Setting the alignment (coaxiality B)
u
Set the magnification of the slit lamp microscope to a low
magnification (e.g. 5x or 6x).
u
Loosen the two clamping screws.
u
Observe the measuring prism through the right eyepiece and
set the correct alignment of the measuring prism. Turn the
tonometer, including the base plate, sideways around the rod.
u
Firmly retighten the two clamping screws alternately.
Adjust alignment
Alignment set
correctly
Adjust alignment
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...