Instruction for use
7 Maintenance
AT 030
7.3 Adjusting the swivel-mounted tonometer holder with round mounting
319300-9120-000-GA-GB-240320
43 / 60
ð
Push the swivel arm including tonometer forwards in the
slot until the sharp focus of the structure visible on one side
is in the middle of the motion range. When moving the
tonometer lift it slightly to avoid it tilting under its own
weight.
ð
Then retighten the screws: first, the two set screws, then
both the clamping screws alternately, starting with the rear
clamping screw.
ð
Carry out a further check in accordance with step 5.
u
If the sharp focus position is only at the rear (towards
observer), the swivel arm with tonometer should be moved
further to the rear:
ð
First loosen the two set screws by approx. ¼ turn.
ð
Then loosen the two clamping screws by approx. ½ turn.
ð
Push the swivel arm including tonometer backwards in the
slot until the sharp focus of the structure visible on one side
is in the middle of the motion range. When moving the
tonometer lift it slightly to avoid it tilting under its own
weight.When moving the tonometer lift it slightly to avoid
it tilting under its own weight.
ð
Then retighten the screws: first, the two set screws, then
both the clamping screws alternately, starting with the rear
clamping screw.
ð
Carry out a further check in accordance with step 5.
NOTE! The measuring prism must be visible in the center of
the eyepiece.
Sharp focus position
front
Sharp focus position
set correctly
Sharp focus position
rear
Setting the lateral middle position in the field of view (lateral
adjustment D)
u
Set the magnification of the slit lamp microscope to 25x.
u
Observe the measuring prism through the microscope (select
proper eyepiece!).
u
If the measuring prism is to the right of the center of the field
of view, proceed as follows:
ð
Loosen the right set screw slightly.
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...