Instruction for use
7 Maintenance
AT 030
7.4 Checking the device
319300-9120-000-GA-GB-240320
45 / 60
Detailed check before mounting on the slit lamp
1
Figure 13: Checking the AT 030
1
Test weight
u
Set the scale reading to 0.
u
Take the 2 g and 5 g test weights out of the storage box.
u
Put the AT 030 onto the open box with the scale facing
downwards.
u
Insert the measuring prism inversely up to the normal
measuring position into the measuring prism holder.
u
Place a test weight concentrically onto the measuring prism.
u
Turn the rotary knob slowly from the 0°-position until the
measuring arm with the weight oscillates weightlessly around
its center position. Remove the test weight.
u
Record the corresponding scale value.
u
Repeat this procedure four times (two times each from the side
of greater and smaller forces) and calculate the arithmetic
average of the readings.
NOTE! The values obtained with both weights should be 2.0
± 0.06 sc dlv. (2 g weight) and 5.0 ± 0.1 sc dlv. (5 g weight).
If these values cannot be reproduced, discontinue using the
tonometer and consult ZEISS Service.
Please also observe the
information in section Metrological check [
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...