Instruction for use
7 Maintenance
AT 030
7.2 Adjusting the swivel-mounted tonometer holder with square mounting
319300-9120-000-GA-GB-240320
37 / 60
If the same eyepiece is always to be used for tonometry, the
adjustment is now finished.
If, however, both eyepieces are to be used alternately, the position
of the AT 030 on the base plate must now be changed:
u
Release the knurled screw and remove the tonometer.
u
Set the tonometer with the centering pin into the other
alignment hole of the base plate.
u
Retighten the knurled screw.
u
Check, whether the measuring prism is also correctly aligned in
this eyepiece. If not, repeat the first four steps.
Setting/checking the correct focus position (image focus C)
u
Set the scale on the tonometer to zero.
u
To assess the focus sharpness apply a visible structure to one
side of the front surface of the measuring prism (e.g. finger-
print, dust, graph paper).
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...