Instruction for use
5 Daily startup
AT 030
5.1 Preparation safety
319300-9120-000-GA-GB-240320
23 / 60
5 Daily startup
5.1 Preparation safety
CAUTION!
Risk due to contaminated or damaged measuring prisms
The patient may be injured by contaminated or damaged
measuring prisms.
u
Never use damaged measuring prisms!
u
Only use clean, residue-free and undamaged measuring prisms.
u
Prior to every use, the contact surface of the measuring prism
must be inspected for contaminants and damage (scratches,
cracks or sharp edges). This must be performed with a slit lamp
microscope at a medium magnification (e.g. 12x or 16x).
CAUTION!
Risk of non-compliance with contraindications
The patient may be injured or suffer complications if the contraindi-
cations are ignored.
u
The device should not be used if the patient is susceptible to
the anesthetic used for the tonometry.
u
Utmost caution is also advised with patients suffering from
corneal epithelium lesions.
u
Utmost caution is also advised with patients with keratitis,
infection of the cornea or penetrating eye injuries.
CAUTION!
Risk due to measuring errors
Measurements with a damaged tonometer may be faulty.
u
Check the tonometer for visible damage as described in the
chapter Checking the tonometer [
Prior to using the device, the user must ensure that it is in a good
condition and fully functioning. Furthermore, the user must follow
the instructions in the user manual.
NOTE! The tonometric measurement should be taken in a
darkened room and the slit lamp operated at normal lamp
intensity. Only under these conditions is it possible to repro-
ducibly observe the weakly shining inner edge of the
fluorescein ring. To ensure that the scale can also be seen in
darkened rooms, it is set against a green fluorescent
background.
NOTE! The actual measuring process should take no longer
than 30 seconds, otherwise the pressure will be reduced by
flowing processes.
Summary of Contents for AT 030
Page 1: ...AT 030 Applanation tonometer Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...AT 030 Applanation tonometer Instruction for use...
Page 8: ......
Page 16: ...Empty page for your notes...
Page 20: ...Empty page for your notes...
Page 28: ...Empty page for your notes...
Page 36: ...Empty page for your notes...
Page 38: ...Empty page for your notes...
Page 52: ...Empty page for your notes...
Page 56: ...Empty page for your notes...
Page 60: ...Empty page for your notes...
Page 62: ...Empty page for your notes...
Page 64: ...Empty page for your notes...
Page 68: ......
Page 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Page 70: ......
Page 72: ...Empty page for your notes...
Page 85: ...Empty page for your notes...
Page 87: ...Empty page for your notes...