146
SECTION:
POSSIBLE CAUSE:
Calibrator bottle not primed.
ACTION:
Prime calibrator bottle.
SECTION:
POSSIBLE CAUSE:
Low calibrator solution.
ACTION:
Install new calibrator.
SECTION:
Install Calibrator Solution(s)
POSSIBLE CAUSE:
Pinched, blocked, leaking or disconnected tube.
ACTION:
Fix or install new tubing.
SECTION:
POSSIBLE CAUSE:
Calibrator pump not operating properly.
ACTION:
Check pump and tubing.
SECTION:
POSSIBLE CAUSE:
Fluid not conductive.
ACTION:
Use saline as diluent.
SYMPTOM:
Fail FCN Test
POSSIBLE CAUSE:
Damaged or old membrane.
ACTION:
Replace membrane.
SECTION:
SYMPTOM:
Fail Linearity Test
POSSIBLE CAUSE:
Probe assignment incorrect.
ACTION:
Make correct assignment.
SECTION:
4.6.1 Assign Chemistries to Probes
POSSIBLE CAUSE:
Calibrator bottle assignment incorrect.
ACTION:
Make correct assignment.
SECTION:
POSSIBLE CAUSE:
Damaged or old Membrane.
ACTION:
Replace membrane.
SECTION:
POSSIBLE CAUSE:
Contaminated or old calibration or linearity standard.
ACTION:
Repeat test with new standards.
SECTION:
5.1 Perform Daily Operational Checks
POSSIBLE CAUSE:
Assigned concentration range beyond practical limits.
ACTION:
Redefine measurement parameters. Remake standards.
SECTION:
SYMPTOM:
ERROR: Motor Failure
POSSIBLE CAUSE:
One of the motors is jammed.
ACTION:
Enter motor service and cycle the suspected motor.
SECTION:
POSSIBLE CAUSE:
Worn sipper pump seals.
ACTION:
Replace seals.
SECTION:
9.4.5 Sipper pump Seal Replacement
SYMPTOM:
ERROR: Temperature
POSSIBLE CAUSE:
Ambient temperature too cold or hot.