5-33
IM XL100C-E
1
2
3
4
5
6
7
App
Index
Standard Protocol
p1
Filter setting
None:
OFF
50 Hz:
50Hz
60 Hz:
60Hz
Query
XI?
Example
Set the filter to 60 Hz.
XI 60HZ
Description In order to activate the settings that are changed
using the basic setting commands, the settings
must be saved using the XE command. If you do
not, the settings before the change are activated
if logging is started.
XB
Burnout setting (query only)
Syntax
XB p1,p2<terminator>
p1
Channel number
For 8 ch input:
01
to
08
For 16 ch input:
01
to
16
p2
Burnout procedure
UP (fixed)
Query
XB[ p1]?
Description Only a query is available for this command.
XJ
RJC setting (query only)
Syntax
XJ p1,p2<terminator>
p1
Channel number
For 8 ch input:
01
to
08
For 16 ch input:
01
to
16
p2
Reference junction compensation selection
INTERNAL
Query
XJ[ p1]?
Example
Only a query is available for this command.
XV
Sets the sampling interval
Syntax
XV p1<terminator>
p1
Sampling Interval
For 8 ch input:
100MS, 200MS, 500MS, 1S, 2S,
5S, 10S, 20S, 30S, 1MIN, 2MIN,
5MIN, 10MIN, 20MIN, 30MIN,
60MIN, 1H
For 16 ch input:
100MS, 200MS, 500MS, 1S, 2S,
5S, 10S, 20S, 30S, 1MIN, 2MIN,
5MIN, 10MIN, 20MIN, 30MIN,
60MIN, 1H
Query
XV?
Example
Set the sampling interval to 1 s.
XV 1S
Description This command is the same as the SW setting
command.
XT
Selects the temperature unit
Syntax
XT p1<terminator>
p1
Temperature unit
Celsius (
°
C):
C
Fahrenheit (
°
F):
F
Query
XT?
Example
Set the temperature unit to Fahrenheit.
XT F
XU
Sets the language
Syntax
XU p1,p2,p3,p4,p5<terminator>
p1
Reserved
p2
Reserved
p3
English:
ENGLISH
Japanese:
JAPANESE
p4
Reserved
p5
Reserved
Query
XU?
Example
Set the language to Japanese.
XU ,,JAPANESE
Description Parameters p1, p2, p4, and p5 are undefined.
XD
Sets the data save operation
Syntax
XD p1,p2,p3,p4 <terminator>
p1
Data save medium
Internal memory:
0
CF card:
1
SD card:
2
p2
Data type
Binary format:
BINARY
ASCII format:
ASCII
p3
File name
Between 1 and 12 characters:
Saves the data to a specified file
name
When omitted:
The file name is automatically
assigned by the XL100.
p4
Printer output
ON: ON
OFF: OFF
Query
XD?
Example
Set the data save destination to the CF card, the
save format to binary, the file name to abcdef,
and the printer output to OFF.
XD 1,BINARY,abcdef,OFF
Description • If the file name of parameter p3 is omitted, the
XL100 is set in a mode in which the file name
is automatically assigned while logging. For
details, see the XL100 User’s Manual.
• In order to activate the settings that are
changed using the basic setting commands,
the settings must be saved using the XE
command. If you do not, the settings before
the change are activated if logging is started.
XO
Selects the communication
method
When switching to Ethernet communications
Syntax
XO p1<terminator>
p1
Ethernet communication
ETHERNET
5.7 Basic Setting Commands