Standard Protocol
5.4
Input Range Parameters
The following tables show which input ranges of the instrument correspond to the
channel modes of the SR command (input range setting command), VOLT, TC, RTD,
pulse, and logic as well as the ranges for the upper and lower limits of the span. Another
table lists the unit groups and units of the SR, SO, and SM commands when the analog
channel scaling is ON.
• DC voltage (VOLT)
Input Range
Range Parameter
Selectable Range of Span
Range of Span Parameter
of the SR Command
Upper and Lower Limits
of the SR Command
of the Input Range
100 mV
100mV
-100.00 to 100.00 mV
-10000 to 10000
500 mV
500mV
-500.0 to 500.0 mV
-5000 to 5000
1 V
1V
-1.0000 to 1.0000 V
-10000 to 10000
5 V
5V
-5.000 to 5.000 V
-5000 to 5000
10 V
10V
-10.000 to 10.000 V
-10000 to 10000
30 V
30V
-30.00 to 30.00 V
-3000 to 3000
1-5 V
1-5V
1.000 to 5.000 V
1000 to 5000
• Thermocouple (TC)
Input Range
Range Parameter
Selectable Range of Span
Range of Span Parameter
of the SR Command
Upper and Lower Limits
of the SR Command
of the Input Range
R
R
0 to 1768
°
C
0 to 1768
S
S
0 to 1768
°
C
0 to 1768
B
B
600 to 1800
°
C
600 to 1800
K
K
-200.0 to 1372.0
°
C
-2000 to 13720
E
E
-200.0 to 1000.0
°
C
-2000 to 10000
J
J
-200.0 to 1200.0
°
C
-2000 to 12000
T
T
-200.0 to 400.0
°
C
-2000 to 4000
N
N
0.0 to 1300.0
°
C
0 to 13000
W
W
0 to 2315
°
C
0 to 2315
L
L
-200.0 to 900.0
°
C
-2000 to 9000
U
U
-200.0 to 400.0
°
C
-2000 to 4000
• RTD
Input Range
Range Parameter
Selectable Range of Span
Range of Span Parameter
of the SR Command
Upper and Lower Limits
of the SR Command
of the Input Range
Pt100
PT
-200.0 to 850.0
°
C
-2000 to 8500
JPt100
JPT
-200.0 to 500.0
°
C
-2000 to 5000
• Logic
Input Range
Range Parameter
Selectable Range of Span
Range of Span Parameter
of the SR Command
Upper and Lower Limits
of the SR Command
of the Input Range
None
NONE
0 or 1
0 or 1