5-5
IM XL100C-E
1
2
3
4
5
6
7
App
Index
Standard Protocol
• Example
EA
SR 01,VOLT,20mV,0,20
SR 02,VOLT,20mV,0,20
......
EN
Decimal Place/Unit Information
• The FE command is used to output the data.
• The measured/computed data that are output using the FD command contains only
the mantissa when binary data output is specified. By combining with the binary
display data or the decimal place obtained with the FE command, the correct
measured/calculated data can be derived.
• Syntax
The data is output for each channel in the following syntax.
EA
<CRLF>
s_kccuuuuuu,pp
<CRLF>
............
EN
<CRLF>
s
Data status (N, S)
N
: Normal
S
: Skip (when the input range is set to SKIP for a measurement channel)
k
Channel type
0
: Measurement channel
P
: Pulse channel
A
: Calculation channel
C
: Communication channel
cc
Channel number
01
to
32
uuuuuu
Unit information (6 characters, left-justified)
mV____:mV / V_____:V / ^C____:
°
C / Mc____:Mc / kc___:kc / rpm__:rpm /
xxxxxx
: (User-defined character string)
pp
Decimal place (00 to 04)
No decimal (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.
_
Space
• Example
EA
N 001mV ,01
N 002mV ,01
EN
5.2 Response Syntax