<11. Inspection and Maintenance>
11-2
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
F11.1E.ai
Calibration gas tube
Exploded view of components
Rod
(with outside diameter
of 2 to 2.5 mm)
Figure 11.1 Cleaning the Calibration Gas Tube
11.1.2
Replacing the Sensor Assembly
The performance of the sensor (cell) deteriorates as its surface becomes soiled during operation.
Therefore, you have to replace the sensor when its life expectancy expires, for example, when
it can no longer satisfy a zero gas ratio of 100 ± 30% or a span gas ratio of 0 ± 18%. In addition,
the sensor assembly is to be replaced if it becomes damaged and can no longer operate during
measurement.
If the sensor becomes no longer operable (for example, due to breakage), investigate the cause
and remedy the problem as much as possible to prevent recurrence.
CAUTION
• If the sensor assembly is to be replaced, allow enough time for the detector to cool down
from its high temperature. Otherwise, you may get burned.
• If the cell assembly is to be replaced, be sure to replace the metal O-ring and the contact
together.
Additionally, even in a case where the cell is not replaced, if the contact becomes deformed
and cannot make complete contact with the cell, replace the contact.
• If there is any corroded or discolored area in the metal O-ring groove in which the contact
is embedded, sand the groove with sandpaper or use a metal brush, and then sand further
with a higher grade of sandpaper (no. 1500 or so), or use an appropriate metal brush
to eliminate any sharp protrusions on the groove. The contact’s resistance should be
minimized.
• Use sensor assemblies manufactured in or after Sept. 2000: the serial number on the side
of the sensor assembly should be 0J000 or later (for example: 0K123, 1AA01 etc)
1. Identifying parts to be replaced
In order not to lose or damage disassembled parts, identify the parts to be replaced from among
all the parts in the sensor assembly. Normally, replace the sensor, metal O-ring and contact
together at the same time. If required, also replace the U-shaped pipe, bolts, filter and associated
spring washers.
2. Removal procedures
(1) Remove the four bolts and associated washers from the tip of the detector probe.
(2) Remove the U-shaped pipe support together with the U-shaped pipe. Remove filter also.
(3) Pull the sensor assembly toward you while turning it clockwise. Also, remove the metal
O-ring between the assembly and the probe. Remove filter also.
(When replacing the assembly, be careful not to scratch or dent the tip of the probe with
which the metal O-ring comes in contact (the surface with which the sensor flange also
comes in contact). Otherwise, the sample gas will not be sealed.)
(4) Use tweezers to pull the contact out of the groove in the tip of the probe.
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...