<8. Detailed Data Setting>
8-3
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
40
50
60
70
80
90
100
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
F8-2E.ai
Minimum humidity (for a 4-mA current), % H
2
O
Maximum humidity (for a 20-mA
current), % H
2
O
Ranges over which oxygen
concentrations can be set
Outside ranges
Figure B
Max. and Min. Humidity Set Ranges
l
“Mixing ratio” setting range
The minimum mixing ratio is set to 0 kg/kg or ranges from 0.201 to 0.625 kg/kg. The maximum
“mixing ratio” setting ranges from 0.2 to 1.0 kg/kg, and must be greater than 1.3 times plus 0.187
the mixing ratio set for the minimum.
Setting example 1
If the setting (for a 4 mA current) is 0 kg/kg, you must set the maximum (20 mA) point at more
than 0.2 kg/kg.
Setting example 2
If the setting (for a 4 mA current) is 0.201 kg/kg, you must set the maximum (20 mA) point at more
than 0.449 kg/kg, (0.201 3 1.3 + 0.187 kg/kg). (Numbers after the decimal point are rounded up.)
0.9
0.85
0.8
0.75
0.7
0.65
0.6
0.55
0.5
0.45
0.4
0.201
0.25
0.3
0.35
0.4
0.45
0.5
0.55
0.6
0.65
Maximum mixing ratio, kg/kg at 20 mA
Minimum mixing ratio, kg/kg at 4 mA
Ranges over which oxygen
concentrations can be set
Outside ranges
1
0.95
F8-3E.ai
Figure C
Max. and Min. Mixing Ratio Set Ranges
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...