<10. Other Functions>
10-23
IM 11M12A01-02E 16th Edition : Mar. 19, 2020-00
Enter
Zero calibration
Close the zero gas valve.
Span calibration
End
Manual calibration
F10.20E.ai
Figure 10.20
Zero-point Calibration Complete (in Manual Calibration)
(3) Close the zero gas flow setting valve to stop the zero gas flow. If the valve has a lock nut,
be sure to tighten the lock nut to prevent the any leakage of the zero gas into the sensor
because the valve may become loose during measurement.
10.8.4
Operation After Calibration
No special operation of the instrument is needed after calibration. However, it is recommended
that the pressure reducing valve for the zero gas cylinders be closed because calibration is not
required so often.