<7. Startup>
7-14
IM 11M12A01-02E
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
Then select “Calibration” in the Execution/Setup display.
Enter
Calibration
Manual calibration
F7.17E.ai
Check the span gas
concentration value
Next
Cancel calibration
Change value: 21.00%
Enter
Manual calibration
F7.18E.ai
Figure 7.28
Calibration Display
Figure 7.29
Manual Calibration
(2) Press the [Enter] key to select “Span calibration”. The Manual calibration display shown
in Figure 7.29 then appears. Check that the oxygen concentration for the span gas in this
display coincides with the oxygen concentration in the calibration gas actually used. If the
check results are assumed to be OK, select “Next” in the Manual calibration display.
(3) Follow the display message in Figure 7.30 to turn on span gas flow. Open the span gas flow
valve for the Flow Setting Unit by loosening the valve lock nut and slowly turning the valve
shaft counterclockwise to flow the span gas at 600 ± 60 ml/min. Use the calibration gas
flowmeter to check the flow.
Enter
Open span gas valve.
Set flow span gas to
600ml/min.
Valve opened
Cancel calibration
Manual calibration
F7.19E.ai
Enter
21.00%
0.5min./div
1.00%
21.0 %O
2
CAL. TIME
Manual calibration
F7.20E.ai
Figure 7.30
Span gas Flow Display
Figure 7.31
Span gas Trend Graph
(for Manual Calibration)
(4) If “Valve opened” is selected as in Figure 7.30, an oxygen concentration trend graph (with
the oxygen concentration being measured) appears (see Figure 7.31). The CAL.TIME in
the bottom area of the panel flashes. Observe the trend graph and wait until the measured
value stabilizes in the vicinity of 21% on the graph. At this point, calibration has not yet been
executed yet, so even if the measured value is above or below 21%, no problem occurs.
(5) After the measured value has stabilized, press the [Enter] key to display the “Span
calibration complete” display shown in Figure 7.32. At this point, the measured value is
corrected to equal the span gas concentration setting. Close the span gas flow valve. The
valve lock nut should be tightened completely so that the span gas does not leak.
16th Edition : Mar. 19, 2020-00
Summary of Contents for EXAxt ZR22G
Page 15: ...Blank Page...
Page 51: ...Blank Page...
Page 77: ...Blank Page...
Page 137: ...Blank Page...
Page 171: ...Blank Page...
Page 195: ...Blank Page...
Page 197: ......
Page 199: ......
Page 201: ......
Page 203: ......
Page 205: ......
Page 207: ......
Page 211: ...Blank Page...